Head and Neck Squamous Cell Carcinoma: Short-Term Repeatability of Apparent Diffusion Coefficient and Intravoxel Incoherent Motion Parameters at 3.0T
Y Ding*, C Fuller , A Mohamed , J Wang , J Hazle , UT MD Anderson Cancer Center, Houston, TX
Presentations
TU-G-CAMPUS-I-11 (Tuesday, July 14, 2015) 5:30 PM - 6:00 PM Room: Exhibit Hall
Purpose: Many published studies have recently demonstrated the potential value of intravoxel incoherent motion (IVIM) analysis for disease evaluation. However, few have questioned its measurement repeatability/reproducibility when applied. The purpose of this study was to determine the short-term measurement repeatability of apparent diffusion coefficient ADC, true diffusion coefficient D, pseudodiffusion coefficient D* and perfusion fraction f, in head and neck squamous cell carcinoma (HNSCC) primary tumors and metastatic nodes.
Methods: Ten patients with known HNSCC were examined twice using echo-planar DW-MRI with 12 b values (0 to 800 s/mm2) 1hour to 24 hours apart before radiation treatment. All patients were scanned with the customized radiation treatment immobilization devices to reduce motion artifacts and to improve image registration in repeat scans. Regions of interests were drawn in primary tumor and metastases node in each patient (Fig. 1). ADC and IVIM parameters D, D* and f were calculated by least squares data fitting. Short-term test–retest repeatability of ADC and IVIM parameters were assessed by measuring Bland–Altman limits of agreements (BA-LA).
Results: Sixteen HNSCC lesions were assessed in 10 patients. Repeatability of perfusion-sensitive parameters, D* and f, in HNSCC lesions was poor (BA-LA: -144% to 88% and -57% to 96% for D* and f, respectively); a lesser extent was observed for the diffusion-sensitive parameters of ADC and D (BA-LA: -34% to 39% and -37% to 40%, for ADC and D, respectively) (Fig. 2).
Conclusion: Poor repeatability of D*/f and good repeatability for ADC/D were observed in HNSCC primary tumors and metastatic nodes. Efforts should be made to improve the measurement repeatability of perfusion-sensitive IVIM parameters.