An Evaluation of Out-Of-Field Doses for Electron Beams From Modern Varian and Elekta Linear Accelerators


C Cardenas

C Cardenas*, P Nitsch , R Kudchadker , R Howell , S Kry , UT MD Anderson Cancer Center, Houston, TX

Presentations

TU-G-CAMPUS-T-4 (Tuesday, July 14, 2015) 4:30 PM - 5:00 PM Room: Exhibit Hall


Purpose: Accurately determining out-of-field doses when using electron beam radiotherapy is of importance when treating pregnant patients or patients with implanted electronic devices. Scattered doses outside of the applicator field in electron beams have not been broadly investigated, especially since manufacturers have taken different approaches in applicator designs.

Methods: In this study, doses outside of the applicator field were measured for electron beams produced by a 10x10 applicator on two Varian 21iXs operating at 6, 9, 12, 16, and 20 MeV, a Varian TrueBeam operating at 6, 9, 12, 16, and 20 MeV, and an Elekta Versa HD operating at 6, 9, 12 and 15 MeV. Peripheral dose profiles and percent depth doses were measured in a Wellhofer water phantom at 100 cm SSD with a Farmer ion chamber. Doses were compared to peripheral photon doses from AAPM’s Task Group #36 report.

Results: Doses were highest for the highest electron energies. Doses typically decreased with increasing distance from the field edge but showed substantial increases over some distance ranges. Substantial dose differences were observed between different accelerators; the Elekta accelerator had much higher doses than any Varian unit examined. Surprisingly, doses were often similar to, and could be much higher than, doses from photon therapy. Doses decreased sharply with depth before becoming nearly constant; the dose was found to decrease to a depth of approximately E(MeV)/4 in cm.

Conclusion: The results of this study indicate that proper shielding may be very important when utilizing electron beams, particularly on a Versa HD, while treating pregnant patients or those with implanted electronic devices. Applying a water equivalent bolus of Emax(MeV)/4 thickness (cm) on the patient would reduce fetal dose drastically for all clinical energies and is a practical solution to manage the potentially high peripheral doses seen from modern electron beams.