Validation of a Determinant Based Photon Transport Solver in Dose Perturbed By Diverse Media

no image available
H Kuo

H Kuo1*, W Tome1 , R Yaparpalvi1 , M Garg1 , W Bodner1 , S Kalnicki1 , (1) Montefiore Medical Center/ALBERT EINSTEIN COLLEGE OF MEDICINE, Bronx, NY

Presentations

SU-E-T-791 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To validate a determinant based photon transport solver in dose imparted within different transition zone between different medium.

Methods:Thickness (.2cm, .5cm, 1cm, 3cm) from various materials (Air -density=0.0012g/cm3, Cork-0.19g/cm3, Lung-0.26g/cm3, Bone-1.85g/cm3, Aluminum (Al)-2.7g/cm3, Titanium (Ti)-4.42g/cm3, Iron (Fe)-8g/cm3) were sandwiched by 10cm solid water. 6MV were used to study the calculation difference between a superposition photon beam model (AAA) and the determinant based Boltzmann photon transport solver (XB) at the upstream (I) and downstream boarder (II) of the medium, within the medium (III), and at far distance downstream away from medium (IV). Calculation was validated with available thickness of Air, Cork, Lung, Al, Ti and Fe. Results are presented as the ratio of the dose at the point with medium perturbation to the same point dose without perturbation.

Results:Zone I showed different backscatter enhancement from high-density materials within the 5mm of the upstream border. AAA showed no backscatter at all, XB showed good agreement beyond 1mm upstream (1.18 vs 1.14, 1.09 vs 1.10, and 1.04 vs 1.05 for Fe, Ti, and Fe, respectively). Zone II showed a re-buildup after exiting high-density medium and Air but no build up for density close to water in both of the measurement and XB. AAA yielded the opposite results in Zone II. XB and AAA showed in Zone III very different absorption in high density medium and the Air. XB and measurement had high concordance regarding photon attenuation in Zone IV. AAA showed less agreement especially when the medium was Air or Fe.

Conclusion:XB compared well with measurement in regions 1mm away from the interface. Planning using XB should be beneficial for External Beam Planning in situations with large air cavity, very low lung density, compact bone, and any kind of metal implant.