Program Information
Eye Lens Dose Reduction From CT Scan Using Organ Based Tube Current Modulation
H Liu1,2*, T Liu2 , X Xu2 , J Wu1 , W Zhuo1 , (1) Fudan University, Shanghai, Shanghai, (2) Rensselaer Polytechnic Inst., Troy, NY
Presentations
SU-E-I-37 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:
To investigate the eye lens dose reduction by CT scan with organ based tube current modulation (OBTCM) using GPU Monte Carlo code ARCHER-CT.
Methods:
36 X-ray sources and bowtie filters were placed around the patient head with the projection angle interval of 10° for one rotation of CT scan, each projection was simulated respectively. The voxel eye models with high resolution(0.1mm*0.1mm*0.1mm) were used in the simulation and different tube voltage including 80kVp, 100kVp, 120kVp and 140kVp were taken into consideration.
Results:
The radiation doses to the eye lens increased with the tube voltage raised from 80kVp to 140kVp, and the dose results from 0° (AP) direction are much higher than those from 180° (PA) direction for all the 4 different tube voltage investigated. This 360° projection dose characteristic enables organ based TCM, which can reduce the eye lens dose by more than 55%.
Conclusion:
As the eye lens belongs to superficial tissues, its radiation dose to external exposure like CT is direction sensitive, and this characteristic feature makes organ based TCM to be an effective way to reduce the eye lens dose, so more clinical use of this technique were recommended.
Funding Support, Disclosures, and Conflict of Interest: National Nature Science Foundation of China(No.11475047)
Contact Email: