Quantitative Assessment of Plan Robustness for Helical Tomotherapy for Head and Neck Cancer Radiotherapy
J Matney1*, J Lian1 , E Chao2 , B Chera1 , L Marks1 , S Das1 , (1) UNC School of Medicine, Chapel Hill, NC, (2) Accuray Incorporated, Madison, WI,
Presentations
SU-E-T-657 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Introduction: Geometric uncertainties in daily patient setup can lead to variations in the planned dose, especially when using highly conformal techniques such as helical Tomotherapy. To account for the potential effect of geometric uncertainty, our clinical practice is to expand critical structures by 3mm expansion into planning risk volumes (PRV). The PRV concept assumes the spatial dose cloud is insensitive to patient positioning. However, no tools currently exist to determine if a Tomotherapy plan is robust to the effects of daily setup variation. We objectively quantified the impact of geometric uncertainties on the 3D doses to critical normal tissues during helical Tomotherapy.
Methods: Using a Matlab-based program created and validated by Accuray (Madison, WI), the planned Tomotherapy delivery sinogram recalculated dose on shifted CT datasets. Ten head and neck patients were selected for analysis. To simulate setup uncertainty, the patient anatomy was shifted ±3mm in the longitudinal, lateral and vertical axes. For each potential shift, the recalculated doses to various critical normal tissues were compared to the doses delivered to the PRV in the original plan
Results: 18 shifted scenarios created from Tomotherapy plans for three patients with head and neck cancers were analyzed. For all simulated setup errors, the maximum doses to the brainstem, spinal cord, parotids and cochlea were no greater than 0.6Gy of the respective original PRV maximum. Despite 3mm setup shifts, the minimum dose delivered to 95% of the CTVs and PTVs were always within 0.4Gy of the original plan.
Conclusions: For head and neck sites treated with Tomotherapy, the use of a 3mm PRV expansion provide a reasonable estimate of the dosimetric effects of 3mm setup uncertainties. Similarly, target coverage appears minimally effected by a 3mm setup uncertainty. Data from a larger number of patients will be presented. Future work will include other anatomical sites.