Use of Electronic Portal Imaging Device (EPID) for Quality Assurance (QA) of Electron Beams On Varian Truebeam System
B Cai*, S Yaddanapudi , B Sun , H Li , C Noel , S Mutic , S Goddu , Department of Radiation Oncology, Washington University in St Louis, St. Louis, MO
Presentations
SU-E-T-775 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:
In a previous study we have demonstrated the feasibility of using EPID to QA electron beam parameters on a single Varian TrueBeam LINAC. This study aims to provide further investigation on (1) reproducibility of using EPID to detect electron beam energy changes on multiple machines and (2) evaluation of appropriate calibration methods to compare results from different EPIDs.
Methods:
Ad-hoc mode electron beam images were acquired in developer mode with XML code. Electron beam data were collected on a total of six machines from four institutions. A custom-designed double-wedge phantom was placed on the EPID detector. Two calibration methods - Pixel Sensitivity Map (PSM) and Large Source-to-Imager Distance Flood Field (LSID-FF) - were used. To test the sensitivity of EPID in detecting energy drifts, Bending Magnet Current (BMC) was detuned to invoke energy changes corresponding to ~±1.5 mm change in R50% of PDD on two machines from two institutions. Percent depth ionization (PDI) curves were then analyzed and compared with the respective baseline images using LSID-FF calibration. For reproducibility testing, open field EPID images and images with a standard testing phantom were collected on multiple machines. Images with and without PSM correction for same energies on different machines were overlaid and compared.
Results:
Two pixel shifts were observed in PDI curve when energy changes exceeded the TG142 tolerance. PSM showed the potential to correct the differences in pixel response of different imagers. With PSM correction, the histogram of images differences obtained from different machines showed narrower distributions than those images without PSM correction.
Conclusion:
EPID is sensitive for electron energy changes and the results are reproducible on different machines. When overlaying images from different machines, PSM showed the ability to partially eliminate the intrinsic variation of various imagers.
Funding Support, Disclosures, and Conflict of Interest: Research Funding from Varian Medical Systems Inc.Dr. Sasa Mutic receives compensation for providing patient safety training services from Varian Medical Systems, the sponsor of this study.