Encrypted login | home

Program Information

Application of TG-119 for Evaluation of Proton Spot Scanning Based Planning and Treatment Delivery


J Saini

J Saini1*, N Cao1 , S Bowen2 , C Bloch2 , T Wong1 , (1) SCCA Proton Therapy, A Procure Center, Seattle, WA, (2) University of Washington, School of Medicine, Seattle, WA

Presentations

SU-E-T-127 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:The clinical test cases presented in AAPM TG-119 are used to evaluate the accuracy of treatment planning and delivery through spot scanning proton beams.

Methods:An IBA spot scanning delivery system has been commissioned to be used with the RayStation treatment planning system. Various test cases provided in TG-119 were used for planning and delivery verification. The CT dataset and structures as provided by TG-119 were imported into a mock patient. The plans were optimized using the multi field optimization (MFO) to achieve the desired goals. The planner was given the flexibility to achieve the given dose-volume goals by creating appropriate objectives and constraints. Beams were delivered to a phantom and measurements were performed at multiple depths using the MatrixxPT detector array. The analyses were performed on beam by beam basis and quantified using the gamma index. A tolerance of 3%/3 mm in 2D was used for gamma index analysis along with dose threshold of 10%.

Results:The clinical goals for targets and critical structures were met or improved for all cases except the C-Shape target with difficult constraints. The minimum gamma index using the 3%/3mm as a criterion is 93.3% for one of the planes measured for C-Shape target. Using 2%/2mm as a criterion, the minimum gamma index drops to 70%. Only Prostate target has all the planes above >90% pass using the 2%/2mm criterion.

Conclusion:The overall accuracy of the treatment planning and delivery is deemed clinically acceptable. The test cases with highly modulated beams can have steep gradients in the dose profiles that can reduce the gamma index pass rate. Gamma analysis based on 3D data may be needed for routine use of 2%/2mm criterion. In addition, improvements in modelling of spot profiles in dose engine may be required for further improving the gamma index pass rate.


Contact Email: