Establishment of CT Diagnostic Reference Levels in Province Nova Scotia


E Tonkopi

E Tonkopi*, M Abdolell , S Duffy , Dalhousie University and Queen Elizabeth II Health Sciences Centre, Halifax, NS

Presentations

SU-E-I-33 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To evaluate patient radiation dose from the most frequently performed CT examinations and to establish provincial diagnostic reference levels (DRLs) as a tool for protocol optimization.

Methods: The study investigated the following CT examinations: head, chest, abdomen/pelvis, and chest/abdomen/pelvis (CAP). Dose data, volume CT dose index (CTDIvol) and dose-length product (DLP), were collected from 15 CT scanners installed during 2004–2014 in 11 hospital sites of Nova Scotia. All scanners had dose modulation options and multislice capability (16–128 detector rows). The sample for each protocol included 15 average size patients (70±20 kg). Provincial DRLs were calculated as the 75th percentile of patient dose distributions. The differences in dose between hospitals were evaluated with a single factor ANOVA statistical test. Generalized linear modeling was used to determine the factors associated with higher radiation dose. A sample of 36 abdominal studies performed on three different scanners was blinded and randomized for an assessment by an experienced radiologist who graded the imaging quality of anatomic structures.

Results: Data for 900 patients were collected. The DRLs were proposed using CTDIvol (mGy) and DLP (mGy*cm) values for CT head (67 and 1049, respectively), chest (12 and 393), abdomen/pelvis (16 and 717), and CAP (14 and 1034). These DRLs were lower than the published national data except for the head CTDIvol. The differences between the means of the dose distributions from each scanner were statistically significant (p<0.05) for all examinations. A very weak correlation was found between the dose and the scanner age or the number of slices with Pearson’s correlation coefficients of 0.011–0.315. The blinded analysis of image quality demonstrated no clinically significant difference except for the noise category.

Conclusion: Provincial DRLs were established for typical CT examinations. The variations in dose between the hospitals suggested a large potential for optimization of examinations.

Funding Support, Disclosures, and Conflict of Interest: Radiology Research Foundation grant