Purpose

Using diagnostic reference levels (DRL) to optimize CT protocols has potential to reduce radiation dose and meet regulatory requirements. However, DRL’s tend to be misconstrued as dose limits, are typically designed for specific patient populations, and are assumed to have acceptable image quality (AIQ) associated with them. To determine the image quality that is associated with established DRL’s for adult abdominal CT studies, a LCD phantom study was employed.

Methods:

A CT phantom (CIRS) containing three columns of 7 spherical targets, ranging from 10mm to 2.4 mm, that are 5, 10, and 20 HU below the background (HUBB) matrix was scanned with a GE HD750 64 slice scanner.

The phantom was scanned at the NEXT 2006 25th CTDIvol of 12 mGy, the NCRP 172 achievable dose (AD) CTDIvol of 17 mGy and 75th CTDIvol of 25 mGy and at the ACR recommended CTDIvol of 25 mGy. It was also scanned at a CTDIvol 20% greater than the AD at 20 mGy and the ACR maximum threshold of 30 mGy.

Results:

At the NEXT 2006 25th percentile CTDIvol of 12 mGy, a 6.3 mm low contrast lesion was detectable in the 20 HUBB; 6.3 mm in the 10 HUBB and 10 mm in the 5 HUBB column.

Increasing the CTDIvol to the NCRP 172 AD of 17 mGy, an additional 4.8 mm lesion was visualized in the 20 HUBB column. At 20 mGy, an additional 4.8 mm lesion was detectable in the 10 HUBB column.
No further lesions were visible between 20 and 30 mGy. However, conspicuity of all lesions increased with each additional step up in CTDI.

Conclusion:

Optimizing radiation dose to achieve AIQ is a critical aspect of any dose optimization committee. Hence, judicious monitoring of radiation exposure to patients has to be balanced with diagnostic image quality.


Contact Email: