Program Information
Track Repeating Algorithm for Proton Therapy Applied to Intensity Modulated Proton Therapy for Head-And-Neck Patients
P Yepes1*, D Mirkovic2 , R Mohan3 , (1) Rice University,Houston,TX 77005(2) U.T M.D. Anderson Cancer Center, Houston, TX, (3) UT MD Anderson Cancer Center, Houston, TX
Presentations
SU-E-T-764 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:
To determine the suitability of fast Monte Carlo techniques for dose calculation in particle therapy based on track-repeating algorithm for Intensity Modulated Proton Therapy, IMPT. The application of this technique will make possible detailed retrospective studies of large cohort of patients, which may lead to a better determination of Relative Biological Effects from the analysis of patient data.
Methods:
A cohort of six head-and-neck patients treated at the University of Texas MD Anderson Cancer Center with IMPT were utilized. The dose distributions were calculated with the standard Treatment Plan System, TPS, MCNPX, GEANT4 and FDC, a fast track-repeating algorithm for proton therapy for the verification and the patient plans. FDC is based on a GEANT4 database of trajectories of protons in a water. The obtained dose distributions were compared to each other utilizing the g-index criteria for 3mm-3% and 2mm-2%, for the maximum spatial and dose differences. The γ-index was calculated for voxels with a dose at least 10% of the maximum delivered dose. Dose Volume Histograms are also calculated for the various dose distributions.
Results:
Good agreement between GEANT4 and FDC is found with less than 1% of the voxels with a γ-index larger than 1 for 2 mm-2%. The agreement between MCNPX with FDC is within the requirements of clinical standards, even though it is slightly worse than the comparison with GEANT4.The comparison with TPS yielded larger differences, what is also to be expected because pencil beam algorithm do not always performed well in highly inhomogeneous areas like head-and-neck.
Conclusion:
The good agreement between a track-repeating algorithm and a full Monte Carlo for a large cohort of patients and a challenging, site like head-and-neck, opens the path to systematic and detailed studies of large cohorts, which may yield better understanding of biological effects.
Contact Email: