Portable Fast Neutron and Photon Dose Meter

no image available
C Miller

C A Miller*, S D Clarke , S A Pozzi , University of Michigan, Ann Arbor, Michigan

Presentations

WE-AB-BRB-11 (Wednesday, July 15, 2015) 7:30 AM - 9:30 AM Room: Ballroom B


Purpose: To develop an instrument for measuring neutron and photon dose rates from mixed fields with a single device.

Methods: Stilbene organic scintillators can be used to detect fast neutrons and photons. Stilbene was used to measure emission from mixed particle sources californium-252 (Cf-252) and plutonium-beryllium (PuBe). Many source detector configurations were used, along with varying amounts of shielding. Collected spectra were analyzed using pulse shape discrimination software, to separate neutron and photon interactions. With a measured light output to energy relationship the pulse height spectrum was converted to energy deposited in the detector. Energy deposited was converted to dose with a variety of standard dose factors, for comparison to current methods. For validation, all measurements and processing was repeated using an EJ-309 liquid scintillator detector. Dose rates were also measured in the same configuration with commercially available dose meters for further validation.

Results: Measurements of dose rates will show agreement across all methods. Higher accuracy of pulse shape discrimination at lower energies with stilbene leads to more accurate measurement of neutron and photon deposited dose. In strong fields of mixed particles discrimination can be performed well at a very low energy threshold. This shows accurate dose measurements over a large range of incident particle energy.

Conclusion: Stilbene shows promise as a material for dose rate measurements due to its strong ability for separating neutrons and photon pulses and agreement with current methods. A dual particle dose meter would simplify methods which are currently limited to the measurement of only one particle type. Future work will investigate the use of a silicon photomultiplier to reduce the size and required voltage of the assembly, for practical use as a handheld survey meter, room monitor, or phantom installation.

Funding Support, Disclosures, and Conflict of Interest: Funding From the United States Department of Energy and the National Nuclear Security Administration.