Encrypted login | home

Program Information

KVCBCT, MVCBCT and MVCT On-Board Imaging Suitability for An Urgent Radiotherapy Treatment Workflow


M Held

M Held*, O Morin , J Pouliot , UC San Francisco, San Francisco, CA

Presentations

TU-AB-303-10 (Tuesday, July 14, 2015) 7:30 AM - 9:30 AM Room: 303


Purpose: A comparison of image quality and dose calculation accuracy to study the suitability of available on-board imaging systems for a new treatment workflow in emergency radiotherapy situations.

Methods: Water and anthropomorphic phantom images were acquired on four different Linac on-board imaging systems, including kVCBCT (Varian TrueBeam and Elekta VersaHD), MVCBCT (Siemens Artiste) and MVCT (Accuray Tomotherapy). Simple treatments of single or opposed beams were planned on the respective kVCT images and copied to all CT images. Image suitability for dose planning was based on the overall mean dose differences and 3D gamma index with 3%/3mm criteria for a prescription of 100 monitor units (MU) and differences in calculated MUs per plan for dose prescriptions to mid-plane.

Results: TrueBeam kVCBCT and Tomotherapy MVCT images produced most accurate dose calculation for all tested cases (γ-index >95%). MVCBCT and VersaHD kVCBCT images resulted in minimum γ-passing rates of 94% and 87%, respectively. MUs calculated from treatment plans prescribed to mid-plane were within differences of 5% relative to kVCT-based plans in all cases. However, VersaHD images showed considerable local image artifacts in the pelvis water and anthropomorphic neck phantom that complicated accurate Hounsfield unit (HU) to electron density conversion, thus causing local dose differences of more than 10% relative to kVCT-based dose distributions.

Conclusion: TrueBeam kVCBCT, MVCBCT and MVCT systems provide image quality that allows for accurate simple treatment plan calculation. Prescription points should be placed away from areas found to cause local dose discrepancies, such as air cavities. Improved image filter settings and HU-to-electron density calibration adjustments may be required for the VersaHD system to obtain an overall accurate dose distribution. To evaluate a system’s overall suitability, its clinical features also require consideration, such as imaging field of view, acquisition time and the ability to instantly export images for treatment planning.


Contact Email: