Encrypted login | home

Program Information

The Use of Electron Dense Materials for Quality Assurance Testing in a Magnetic Field

no image available
H van Zijp

H van Zijp*, B van Asselen , W de Vries , K Ishakoglu , E Beld , J Kok , J Wolthaus , J Lagendijk , B Raaymakers , University Medical Center Utrecht, Utrecht, Netherlands

Presentations

SU-E-T-745 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: Due to the effect of the magnetic field on the dose deposition (skewing the beam), machine quality assurance (QA) tests for the MR-linac (MRI combined with a linear accelerator) need to be redesigned. In this work we focus on the redesign of QA tests that address geometrical accuracy of the system.

Methods: Using electron dense materials (e.g. copper in our experiment) the dose kernel is minimized and thereby the effect of the magnetic field on the dose distribution. This approach is supported by Monte-Carlo simulations and can be used in practice with film measurements. Two examples of QA tests are presented: beam profile and star-shot measurements.

Results: The novel method was verified by performing both measurements on a conventional linac and the MR-linac with a film that was sandwiched between copper layers. Measurements were compared with a reference setup which was similar to setup used in clinical practice. On a conventional linac the experimental outcome showed good agreement between the reference and the new setup for both QA tests. The results from the MR-linac showed that the symmetry of the beam profile was restored in presence of the copper layers in the setup and that the isocenter size can be determined accurately with the introduced star-shot setup (see supporting material).

Conclusion: The use of electron dense materials for QA tests was shown to be a simple and effective method to remove the effects on the dose distribution enabling assessment of geometrical accuracy of a MR-linac system. The use of high dense materials is not limited to the presented QA tests only, but has a broad applicability for beam specific QA tests in presence of a magnetic field.


Contact Email: