Encrypted login | home

Program Information

Accuracy of Absolute Three-Dimensional Dose Distribution Measurement Using the Delta4

no image available
R Uehara

R Uehara*, H Tachibana , H Ohyoshi , S Matsumoto , H Baba , F Tanaka , T Ariji , National Cancer Center Hospital East, Kashiwa, Chiba

Presentations

SU-E-T-89 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: In this study, we investigated the accuracy of the absolute dose distribution measurement using the Delta4 phantom compared to the measurements using a ionization chamber and EDR2 film
Methods: Several conventional and intensity-modulated radiation therapy plans were used to compare the dose distribution measured using the Delta4 phantom to the absolute point dose using the chamber and the relative two-dimensional dose distribution using the EDR2 film. For the absolute dose distribution evaluation, the measurements using the Delta4, the chamber and the film were performed in similar measurement geometry. For point dose measurement using the chamber, an acrylic slab phantom with the PTW Semiflex chamber was inserted into the Delta4 phantom, alternative to the Delta4 main unit. Similarly, for dose distribution measurement using the film, the EDR2 film sandwiched with two acrylic slab phantoms were inserted to the phantom. Dose difference and gamma analysis were done for point dose and relative dose distribution comparisons, respectively.
Result: The point dose measurements show slight negative systematic dose difference of -0.5 ± 0.1% and -1.0 ± 0.4% in the conventional and the IMRT plans, respectively. The additional measurement for direction dependency for Delta4 shows similar negative systematic dose difference even the phantom analysis software consider the directional dependency. The pass rate of the gamma evaluation was 77.7 ± 5.8% and 88.8±3.3% in the conventional and the IMRT plans, respectively.
Conclusions: The Delta4 phantom shows a 1%-systematic dose difference derived from directional dependency and lower resolution compared to the film. Thus it is necessary to comprehensively evaluate the phantom to verify the IMRT/VMAT plans. Especially, the dosimetry tool is needed to have high resolution and high measurement accuracy in IMRT/VMAT-SBRT plan with small fields using intensity modulation in which the analysis area is limited and the higher correspondence between calculation and measurement is needed.


Contact Email: