Plan Quality Estimation of Intensity-Modulated Radiotherapy Cases for Lung Cancer


J Koo

J Koo1*, M Yoon1 , W Chung2 , D Kim2 , (1) Korea University, Seoul (2) Kyung Hee University Hospital at Gangdong, Seoul

Presentations

SU-E-T-617 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To estimate the planning quality of intensity-modulated radiotherapy in lung cancer cases and to provide preliminary data for the development of a planning quality assurance algorithm.

Methods: 42 IMRT plans previously used in cases of solitary lung cancers were collected. Organs in or near the thoracic cavity, such as lung (ipsilateral, contralateral), heart, liver, esophagus, cord and bronchus were considered as organs at risk (OARs) in this study. The coverage index (CVI), conformity index (CI), homogeneity index (HI), volume, irregularity (standard deviation of center-surface distance) were used to compare PTV dose characteristics. The effective uniform dose (EUD), V10Gy, and V20Gy of the OARs were used to compare OAR dose characteristics.

Results: Average CVI, CI, HI values were 0.9, 0.8, 0.1, respectively. CVI and CI had narrow Gaussian distribution curves without a singular value, but one case had a relatively high (0.25) HI because of location and irregular shape (Irregularity of 18.5 when average was 12.5) of PTV. EUDs tended to decrease as OAR-PTV distance increased and OAR-PTV overlap volume decreased.

Conclusion: This work indicates the potential for significant plan quality deviation of similar lung cancer cases. Considering that this study were from a single department, differences in the treatment results for a given patient would be much more pronounced if multiple departments (and therefore more planners) were involved. Therefore, further examination of QA protocols is needed to reduce deviations in radiation treatment planning.