Dose Build Up and Bolusing Characteristics for Total Body Irradiation Dosimetry
M Butson*, D Pope , M Whitaker , Chris O'Brien LifeHouse, Sydney, NSW
Presentations
SU-E-T-275 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose : Total Body Irradiation (TBI) treatments are mainly used in a preparative regimen for haematopoietic stem cell (or bone marrow) transplantation. Our standard regimen is a 12 Gy / 6 fraction bi-daily technique. To evaluate the delivered dose homogeneity to the patient, EBT3 Gafchromic film is positioned at the head, neck, chest, pelvis and groin for all fractions. This work investigates and quantifies the build-up dose characteristics at TBI distances and requirements for in-vivo dosimetry bolusing.
Methods : Percentage dose build up characteristics of photon beams have been investigated at large extended SSD’s using parallel plate ionisations chambers (Attix) and EBT3 Gafchromic film. Measurements were made to open fields at different field sizes as well as large 40cm x 40cm fields with differing scatter conditions such as the introduction of standard Perspex scattering plates at different distances to the measurement point.
Results: Percentage surface dose measured values for open fields at 300 cm SSD were found to range from 20 % up to 65.5 % for fields of 5 cm x 5 cm to 40 cm x 40 cm. With the introduction of 1cm Perspex scattering plates used in TBI treatments the surface dose values increased up to 83% to 90%, depending on the position of the Perspex scattering plate compared to the measurement point. Our work showed that at least 3mm water equivalent bolus / scatter material should be placed over the EBT3 for accurate dose assessment for TBI treatments.
Conclusion : Build up dose characteristics exist at long (300cm) SSD’s including treatments using Perspex scattering plates placed at various distances form the patient during TBI treatment. Top accurately assess the applied dose during treatment, in-vivo dosimeters such as Gafchromic EBT3 should have at least 3mm bolus / scatter material placed over them to measure actual applied doses.