Investigation of Conformal Arc Therapy Utilizing Newly Designed Cobalt 60 Machine
A Eldib1,2*, O Chibani1 , L Jin1 , J Li1 , I Veltchev1 , R Price1 , G Mora3 , C Ma1 , (1) Fox Chase Cancer Center, Philadelphia, PA, (2) Department of Physics, AlAzhar University, Cairo, Egypt (3) Universidade de Lisboa, Codex, Lisboa
Presentations
SU-C-BRB-5 (Sunday, July 12, 2015) 1:00 PM - 1:55 PM Room: Ballroom B
Purpose:
A new modernized design for cobalt 60 (Co-60) machines is being developed with a ring type gantry. In this study we investigate the beneficial outcome of the new design for conformal arc therapy for various clinical sites. The new modality was evaluated based on isodose distributions and dose volume histograms as compared to 6MV photon beams from conventional linear accelerators.
Methods:
Computed tomographic images of seven different SBRT patients were selected from our patient database. All of these cases were previously planned on the Eclipse treatment planning system. New plans for these patients were generated with a modified conformal arc technique using both 6MV and Co-60 beams. The conformal arc was created by the delivery of treatment fields conformal to the target cross-section at every 5 or 10 degrees. The field shape was modified or turned off when it initially passed through a critical structure. Monte Carlo codes, MCBEAM and MCPLAN, were used for the machine head simulation and phantom/patient dose calculation, respectively. In the new Co-60 machine design, the source-to-isocenter distance was 60cm and the treatment head included the Co-60 source, primary collimator, jaws and MLC.
Results:
For all cases investigated, conformal arc plans utilizing Co-60 beams achieved similar conformity (mean conformity index=1.19) comparing to 6MV photon beams. Isodose distributions were tailored similarly around the PTV; both Co-60 and 6MV plans met our clinical acceptance criteria for the target coverage, and the maximum and minimum target doses. The DVH for the Co-60 plans showed slightly lower doses to the critical structures although the differences were small in most cases.
Conclusion:
There were no significant dosimetric differences between 6MV and Co-60 plans. Our results confirmed that this new Co-60 design could be a cost-effective machine for advanced radiotherapy due to its low cost, low maintenance and high up time.