Utilization of Optical Dosimeter for Modulated Spot-Scanning Particle Beam

no image available
W Hsi

W Hsi*, Y Li , Z Huang , Y SHENG , Y Deng , J Zhao , F Zhao , L Sun , and M Moyers , Shanghai Proton and Heavy Ion Center,Pudong, Shangai, China

Presentations

MO-FG-CAMPUS-T-14 (Monday, July 13, 2015) 5:30 PM - 6:00 PM Room: Exhibit Hall


Purpose: To present the utilization of an optical dosimeter for modulated spot-scanning carbon-ion and proton beams during the acceptance test of Siemens IONTRIS system.

Method and Materials: An optical dosimeter using phosphor scintillation was developed to map and interactively analyze the shapes and sizes of spots over 190 energies for ProTom modulated-scanning system. The dose response to proton had been characterized with proper pixel calibration at ProTom system. The dose response was further studied at 0.7 cm depths by uniform 8cm in-diameter fields of 424.89 MeV/u (E290) carbon-ions and 215.18MeV (E282) protons at IONTRIS system. The virtual source axial distances (vSAD) of carbon-ions and protons of IONTRIS system was investigated by measuring either variations of spot position or field size at five different locations to Isocenter. By measuring lateral profiles of uniform doses with varied thin-thicknesses of chest-board pattern and placing the scintillation plate at near to the distal edge, range variations at different off-axis-distances (rOAD) were examined. Relative accuracy and reproducibility of beam range were measured for three beam ranges with a ramping block at front of scintillation plate.

Results: Similar dose response was observed for high energies of carbon ions and protons. Mean vSAD at X and Y axes were 744.1 cm and 807.4cm with deviation of 7.4cm and 7.7cm, respectively. Variation of rOAD was within 0.35 mm over 10cm for both protons and carbon ions. Accuracy of measuring relative distal range using the ramping block was 0.2mm. Measured range over repeated three times for each range were within 0.25mm at same room, and within 1.0mm between four rooms.

Conclusions: The optical dosimeter could efficiently measure the virtual source distance. And, to measure small range variation at different off-axial locations, and for the relative beam range between rooms during acceptance test of a modulated spot-scanning particle system.