A Procedure for the Fast Semi-Automatic Localization of Catheters Using An Electromagnetic Tracker (EMT) for Image-Guided Brachytherapy

no image available
A Damato

A Damato*, A Viswanathan , R Cormack , Brigham and Women's Hospital, Boston, MA

Presentations

SU-F-BRA-1 (Sunday, July 12, 2015) 4:00 PM - 6:00 PM Room: Ballroom A


Purpose:
To evaluate the feasibility of brachytherapy catheter localization through use of an EMT and 3D image set.

Methods:
A 15-catheter phantom mimicking an interstitial implantation was built and CT-scanned. Baseline catheter reconstruction was performed manually. An EMT was used to acquire the catheter coordinates in the EMT frame of reference. N user-identified catheter tips, without catheter number associations, were used to establish registration with the CT frame of reference. Two algorithms were investigated: brute-force registration (BFR), in which all possible permutation of N identified tips with the EMT tips were evaluated; and signature-based registration (SBR), in which a distance matrix was used to generate a list of matching signatures describing possible N-point matches with the registration points. Digitization error (average of the distance between corresponding EMT and baseline dwell positions; average, standard deviation, and worst-case scenario over all possible registration-point selections) and algorithm inefficiency (maximum number of rigid registrations required to find the matching fusion for all possible selections of registration points) were calculated.

Results:
Digitization errors on average <2 mm were observed for N ≥5, with standard deviation <2 mm for N ≥6, and worst-case scenario error <2 mm for N ≥11. Algorithm inefficiencies were: N = 5, 32,760 (BFR) and 9900 (SBR); N = 6, 360,360 (BFR) and 21,660 (SBR); N = 11, 5.45*1010 (BFR) and 12 (SBR).

Conclusion:
A procedure was proposed for catheter reconstruction using EMT and only requiring user identification of catheter tips without catheter localization. Digitization errors <2 mm were observed on average with 5 or more registration points, and in any scenario with 11 or more points. Inefficiency for N = 11 was 9 orders of magnitude lower for SBR than for BFR.