New Efficient Method for Xoft Axxent Electronic Brachytherapy Source Calibration by Pre-Characterizing Surface Applicators


S Pai

S Pai1*, (1) iCAD Inc., Los Gatos, CA

Presentations

SU-F-BRA-9 (Sunday, July 12, 2015) 4:00 PM - 6:00 PM Room: Ballroom A


Purpose:The objective is to improve the efficiency and efficacy of Xoft™ Axxent™ electronic brachytherapy (EBT) calibration of the source & surface applicator using AAPM TG-61 formalism.

Methods:Current method of Xoft EBT source calibration involves determination of absolute dose rate of the source in each of the four conical surface applicators using in-air chamber measurements & TG61 formalism. We propose a simplified TG-61 calibration methodology involving initial characterization of surface cone applicators. This is accomplished by calibrating dose rates for all 4 surface applicator sets (for 10 sources) which establishes the "applicator output ratios" with respect to the selected reference applicator (20 mm applicator). After the initial time, Xoft™ Axxent™ source TG61 Calibration is carried out only in the reference applicator. Using the established applicator output ratios, dose rates for other applicators will be calculated.

Results:200 sources & 8 surface applicator sets were calibrated cumulatively using a Standard Imaging A20 ion-chamber in accordance with manufacturer-recommended protocols. Dose rates of 10, 20, 35 & 50mm applicators were normalized to the reference (20mm) applicator. The data in Figure 1 indicates that the normalized dose rate variation for each applicator for all 200 sources is better than ±3%. The average output ratios are 1.11, 1.02 and 0.49 for the 10 mm,35 mm and 50 mm applicators, respectively, which are in good agreement with the manufacturer’s published output ratios of 1.13, 1.02 and 0.49.

Conclusion:Our measurements successfully demonstrate the accuracy of a new calibration method using a single surface applicator for Xoft EBT sources and deriving the dose rates of other applicators. The accuracy of the calibration is improved as this method minimizes the source position variation inside the applicator during individual source calibrations. The new method significantly reduces the calibration time to less than one third of the current calibration time while improving efficacy.