Efficient Segmentation of Flattening Filter Free Photon Beamsfor 3D-Conformal SBRT Treatment Planning
J Barbiere*, G Beninati , A Ndlovu , HUMC, Hackensack, New Jersey
Presentations
SU-E-T-356 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose: It has been argued that a 3D-conformal technique (3DCRT) is suitable for SBRT due to its simplicity for non-coplanar planning and delivery. It has also been hypothesized that a high dose delivered in a short time can enhance indirect cell death due to vascular damage as well as limiting intrafraction motion. Flattening Filter Free (FFF) photon beams are ideal for high dose rate treatment but their conical profiles are not ideal for 3DCRT. The purpose of our work is to present a method to efficiently segment an FFF beam for standard 3DCRT planning.
Methods: A 10x10 cm Varian True Beam 6X FFF beam profile was analyzed using segmentation theory to determine the optimum segmentation intensity required to create an 8 cm uniform dose profile. Two segments were automatically created in sequence with a Varian Eclipse treatment planning system by converting isodoses corresponding to the calculated segmentation intensity to contours and applying the “fit and shield” tool. All segments were then added to the FFF beam to create a single merged field. Field blocking can be incorporated but was not used for clarity.
Results: Calculation of the segmentation intensity using an algorithm originally proposed by Xia and Verhey indicated that each segment should extend to the 92% isodose. The original FFF beam with 100% at the isocenter at a depth of 10 cm was reduced to 80% at 4cm from the isocenter; the segmented beam had +/- 2.5 % uniformity up to 4.4cm from the isocenter. An additional benefit of our method is a 50% decrease in the 80%-20% penumbra of 0.6cm compared to 1.2cm in the original FFF beam.
Conclusion: Creation of two optimum segments can flatten a FFF beam and also reduce its penumbra for clinical 3DCRT SBRT treatment.