Purpose: To follow the Integrating Healthcare Enterprise - Radiation Oncology (IHE-RO) initiative of proper cross-vendor technology integration, an automated chart checker (ACC) was developed. ACC compares extracted data from an approved patient plan in the Eclipse treatment planning system (TPS) against data existing in the Mosaiq treatment management system (TMS). ACC automatically analyzes these parameters using built-in quality checklists to provide further aid in chart review.

Methods: Eclipse TPS data are obtained using Eclipse scripting API (ESAPI) while Mosaiq TMS data are obtained from a radiotherapy-treatment-planning (RTP) file. Using this information, ACC identifies TPS-TMS discrepancies in 18 primary beam parameters including MU, energy, jaw positions, gantry angle, table angle, accessories, and bolus for up to 31 beams. Next, approximately 40 items from traditional quality checklists are evaluated such as prescription consistency, DRR graticule placement, plan approval status, global max dose, and dose tracking coefficients. Parameters were artificially modified to determine if ACC would detect an error in data transfer and to test each component of quality checklists.

Results: Using ESAPI scripting and RTP file-processing, ACC was able to properly aggregate data from TPS and TMS for up to 31 beams. Errors were artificially introduced into each plan parameter, and ACC was able to successfully detect all of them within seconds. Next, ACC was able to successfully detect mistakes in the chart by identifying deviations with its quality checklists, within seconds.

Conclusion: ACC effectively addresses the potential issue of faulty cross-vendor data transfer, as described by IHE-RO. In addition, ACC was also able to detect deviations from its built-in quality checklists. ACC is already an invaluable tool for efficient and standardized chart review and will continue to improve as its incorporated checklists become more comprehensive.