IGRT Gently: Evaluating Imaging Dose in Phantoms of Different Sizes


B Morris

B Morris*, W Duggar , J Stanford , C Yang , University of Mississippi Med. Center, Jackson, MS

Presentations

SU-E-J-7 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
IGRT imaging procedures have emerged as a common method of patient position verification in radiotherapy, though imaging dose is generally neglected in the treatment plan. Consequently, evaluating and optimizing the dose from these procedures is worthwhile. This process is especially important for children, who are more radiosensitive than adults. The aim of this work was to gain some understanding of the relative doses involved with various XVI-preset parameters for an “adult” and “child” phantom set, with the hopes that imaging dose for a child can be reduced.

Methods:
32 and 16cm CTDI-phantoms were used as surrogates for adult and child torsos, respectively. Dose was measured in the central and peripheral chamber positions of the phantoms. CBCT scans were made for both phantoms using Elekta’s Chest-preset to establish a dose baseline. The child-phantom was then scanned using the Elekta Head and Neck (HN) preset. A modified HN-preset (named Peds Abd-pelvis) was also created with a doubled mAs to maintain a reduction in dose to the child-phantom (relative to the baseline), while providing clinically-usable image quality.

Results:
The baseline dose to the child-phantom from the Chest-preset was 310% that of the adult-phantom for the center chamber position and 150% at the periphery. An average dose reduction of 97% was obtained in the child-phantom by switching from the Chest-preset to the HN-preset, while the Peds Abd-pelvis-preset similarly reduced the dose by an average of 92%.

Conclusion:
XVI-preset parameters significantly affect dose, and should be optimized to reduce dose, while ensuring clinically-usable image quality.
Using a modified imaging preset (Peds Abd-pelvis-preset) greatly reduced the dose to the child-phantom compared to the dose for the Chest-preset for both the child and adult-phantoms. This outcome provides support for the development of child-specific protocols for IGRT imaging in pediatric patients.