Encrypted login | home

Program Information

ProteusONE Machine QA Procedure and Stabiity Study: Half Year Clinical Operation

no image available
D Freund

D Freund*, X Ding , H Wu , J Zhang , J Syh , J Syh , B Patel , X Song , Willis-Knighton Medical Center, Shreveport, LA

Presentations

SU-E-T-636 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: The objective of this study is to evaluate the stability of ProteusOne, the 1st commercial PBS proton system, throughout the daily QA and monthly over 6 month clinical operation.

Method: Daily QA test includes IGRT position/repositioning, output in the middle of SOBP, beam flatness, symmetry, inplane and crossplane dimensions as well as energy range check. Daily range shifter QA consist of output, symmetry and field size checks to make sure its integrity. In 30 mins Daily QA test, all the measurements are performed using the MatriXXPT (IBA dosimetry). The data from these measurement was collected and compare over the first 6 month of clinical operation. In addition to the items check in daily QA, the summary also includes the monthly QA gantry star shots, absolute position check using a novel device, XRV-100.

Results: Average machine output at the center of the spread out bragg peak was 197.5±.8 cGy and was within 1%of the baseline at 198.4 cGy. Beam flatness was within 1% cross plane with an average of 0.67±0.12% and 2% in-plane with an average of 1.08±0.17% compared to baseline measurements of 0.6 and 1.03, respectively. In all cases the radiation isocenter shift was less than or equal to 1mm. Output for the range shifter was within 2% for each individual measurement and averaged 34.4±.2cGy compare to a baseline reading of 34.5cGy. The average range shifter in and cross plane field size measurements were 19.8±0.5cm and 20.5±0.4cm compared with baseline values of 20.19cm and 20.79cm, respectively. Range shifter field symmetry had an average of less 1% for both in-plane and cross plane measurements.

Conclusion: All machine metrics over the past 6 months have proved to be stable. Although, some averages are outside the baseline measurement they are within 1% tolerance and the deviation across all measurements is minimal.



Contact Email: