Feasibility of a VMAT-Based Spatially Fractionated Grid Therapy Technique

no image available
B Zhao

B Zhao1*, J Jin2 , M Liu1 , Y Huang1 , J Kim1 , S Brown1 , F Siddiqui1 , I Chetty1 , N Wen1 , (1) Henry Ford Health System, Detroit, MI, (2) Georgia Regents University, Augusta, GA

Presentations

TU-CD-304-8 (Tuesday, July 14, 2015) 10:15 AM - 12:15 PM Room: 304


Purpose:
Grid therapy (GT) uses spatially modulated radiation doses to treat large tumors without significant toxicities. Incorporating 3D conformal-RT or IMRT improved single-field GT by reducing dose to normal tissues spatially through the use of multiple fields. The feasibility of a MLC-based, inverse-planned multi-field GT technique has been demonstrated. Volumetric modulated arc therapy (VMAT) provides conformal dose distributions with the additional potential advantage of reduced treatment times. In this study, we characterize a new VMAT-based GT (VMAT-GT) technique with respect to its deliverability and dosimetric accuracy.

Methods:
A lattice of 5mm-diameter spheres was created as the boost volume within a large treatment target. A simultaneous boost VMAT (RapidArc) plan with 8Gy to the target and 20Gy to the boost volume was generated using the Eclipse treatment planning system (AAA-v11). The linac utilized HD120 MLC and 6MV flattening-filter free beam. Four non-coplanar arcs, with couch angles at 0, 45, 90 and 317° were used. Collimator angles were at 45 and 315°. The plan was mapped to a phantom. Calibrated Gafchromic EBT3 films were used to measure the delivered dose.

Results:
The VMAT plan generated a highly spatially modulated dose distribution in the target. D95%, D50%, D5% for the spheres and the targets in Gy were 18.9, 20.6, 23 and 8.0, 9.6, 14.8, respectively. D50% for a 1cm ring 1cm outside the target was 3.0Gy. The peak-to-valley ratio of this technique is comparable to previously proposed techniques, but the MUs were reduced by almost 50%. Film dosimetry showed good agreement between calculated and delivered dose, with an overall gamma passing rate of >98% (3% and 1mm). The point dose differences at sphere centers varied from 2-8%.

Conclusion:
The deliverability and dose calculation accuracy of the proposed VMAT-GT technique demonstrates that ablative radiation doses are deliverable to large tumors safely and efficiently.