Evaluation of a Novel Dedicated Breast PET System (Mammi-PET)
Z Long*, T Swanson , M O'Connor , Mayo Clinic, Rochester, MN
Presentations
WE-AB-204-10 (Wednesday, July 15, 2015) 7:30 AM - 9:30 AM Room: 204
Purpose:To evaluate the performance characteristics of a novel dedicated breast PET system (Mammi-PET, Oncovision). The system has 2 detector rings giving axial/transaxial field of view of 8/17 cm. Each ring consists of 12 monolithic LYSO modules coupled to PSPMTs.
Methods:Uniformity, sensitivity, energy and spatial resolution were measured according to NEMA standards. Count rate performance was investigated using a source of F-18 (1384uCi) decayed over 5 half-lives. A prototype PET phantom was imaged for 20 min to evaluate image quality, recovery coefficients and partial volume effects. Under an IRB-approved protocol, 11 patients who just underwent whole body PET/CT exams were imaged prone with the breast pendulant at 5-10 minutes/breast. Image quality was assessed with and without scatter/attenuation correction and using different reconstruction algorithms.
Results:Integral/differential uniformity were 9.8%/6.0% respectively. System sensitivity was 2.3% on axis, 2.2% and 2.8% at 3.8 cm and 7.8 cm off-axis. Mean energy resolution of all modules was 23.3%. Spatial resolution (FWHM) was 1.82 mm and 2.90 mm on axis and 5.8 cm off axis. Three cylinders (14 mm diameter) in the PET phantom were filled with activity concentration ratios of 4:1, 3:1, and 2:1 relative to the background. Measured cylinder to background ratios were 2.6, 1.8 and 1.5 (without corrections) and 3.6, 2.3 and 1.5 (with attenuation/scatter correction). Five cylinders (14, 10, 6, 4 and 2 mm diameter) each with an activity ratio of 4:1 were measured and showed recovery coefficients of 1, 0.66, 0.45, 0.18 and 0.18 (without corrections), and 1, 0.53, 0.30, 0.13 and 0 (with attenuation/scatter correction). Optimal phantom image quality was obtained with 3D MLEM algorithm, >20 iterations and without attenuation/scatter correction.
Conclusion:The MAMMI system demonstrated good performance characteristics. Further work is needed to determine the optimal reconstruction parameters for qualitative and quantitative applications.