Hiigh Quality and Sub-MSv Cerebral CT Perfusion Imaging


K Li

Ke Li*, Kai Niu , Yijing Wu, Guang-Hong Chen, University of Wisconsin, Madison, WI

Presentations

TU-EF-204-2 (Tuesday, July 14, 2015) 1:45 PM - 3:45 PM Room: 204


Purpose: CT Perfusion (CTP) imaging is of great importance in acute ischemic stroke management due to its potential to detect hypoperfused yet salvageable tissue and distinguish it from definitely unsalvageable tissue. However, current CTP imaging suffers from poor image quality and high radiation dose (up to 5 mSv). The purpose of this work was to demonstrate that technical innovations such as Prior Image Constrained Compressed Sensing (PICCS) have the potential to address these challenges and achieve high quality and sub-mSv CTP imaging.

Methods: (1) A spatial-temporal 4D cascaded system model was developed to indentify the bottlenecks in the current CTP technology; (2) A task-based framework was developed to optimize the CTP system parameters; (3) Guided by (1) and (2), PICCS was customized for the reconstruction of CTP source images. Digital anthropomorphic perfusion phantoms, animal studies, and preliminary human subject studies were used to validate and evaluate the potentials of using these innovations to advance the CTP technology.

Results: The 4D cascaded model was validated in both phantom and canine stroke models. Based upon this cascaded model, it has been discovered that, as long as the spatial resolution and noise properties of the 4D source CT images are given, the 3D MTF and NPS of the final CTP maps can be analytically derived for a given set of processing methods and parameters. The cascaded model analysis also identified that the most critical technical factor in CTP is how to acquire and reconstruct high quality source images; it has very little to do with the denoising techniques often used after parametric perfusion calculations. This explained why PICCS resulted in a five-fold dose reduction or substantial improvement in image quality.

Conclusion: Technical innovations generated promising results towards achieving high quality and sub-mSv CTP imaging for reliable and safe assessment of acute ischemic strokes.

Funding Support, Disclosures, and Conflict of Interest: K. Li, K. Niu, Y. Wu: Nothing to disclose. G.-H. Chen: Research funded, GE Healthcare; Research funded, Siemens AX.