Feasibility for Using Iodine Quantification to Assist Diagnosis in Dual Energy Contrast-Enhanced Digital Mammography
Y Hwang1*, Y Lin2 , Y Cheung1,3 , C Tsai2 , (1) Chang Gung Memorial Hospital at Linkou, Taoyuan,(2) Chang Gung University, Taoyuan, (3) Chang Gung Memorial Hospital at Taoyuan, Taoyuan
Presentations
SU-E-I-45 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:
The objective of this study is to develop quantitative calibration between image quality indexes and iodine concentration with dual-energy (DE) contrast-enhanced digital mammography (CEDM) techniques and further serve as the assistance for diagnosis.
Methods:
Custom-made acrylic phantom with dimensions of 24x30 cm² simulated breast thickness from 2 to 6 cm was used in the calibration. The phantom contained matrix of four times four holes of 3 mm deep with a diameter of 15 mm for filling contrast agent with area density ranged from 0.1 to 10 mg/cm². All the image acquisitions were performed on a full-field digital mammography system (Senographe Essential, GE) with dual energy acquisitions. Mean pixel value (MPV), and contrast-to-noise ratio (CNR) were used for evaluating the relationship between image quality indexes and iodine concentration. Iodine map and CNR map could further be constructed with these calibration curves applied pixel by pixel utilized MATLAB software. Minimum iodine concentration could also be calculated with the visibility threshold of CNR=5 according the Rose model .
Results:
When evaluating the DE subtraction images, MPV increased linearly as the iodine concentration increased with all the phantom thickness surveyed (R² between 0.989 and 0.992). Lesions with increased iodine uptake could thus be enhanced in the color-encoded iodine maps, and the mean iodine concentration could be obtained through the ROI measurements. As for investigating CNR performance, linear relationships were also shown between the iodine concentration and CNR (R² between 0.983 and 0.990). Minimum iodine area density of 1.45, 1.73, 1.80, 1.73 and 1.72 mg/cm² for phantom thickness of 2, 3, 4, 5, 6 cm were calculated based on Rose’s visualization criteria.
Conclusion:
Quantitative calibration between image quality indexes and iodine concentrations may further serving as the assistance for analyzing contrast enhancement for patient participating the dual energy CEDM procedures.