End-To-End (E2E) Testing On TomoHDA System Using a Real Pig Head for Intracranial Radiosurgery

no image available
M Leick

N Corradini , M Leick*, M Bonetti , L Negretti , Clinica Luganese, Radiotherapy Center, Lugano, Switzerland

Presentations

SU-E-J-25 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To determine the MVCT imaging uncertainty on the TomoHDA system for intracranial radiosurgery treatments. To determine the end-to-end (E2E) overall accuracy of the TomoHDA system for intracranial radiosurgery.

Methods: A pig head was obtained from the butcher, cut coronally through the brain, and preserved in formaldehyde. The base of the head was fixed to a positioning plate allowing precise movement, i.e. translation and rotation, in all 6 axes. A repeatability test was performed on the pig head to determine uncertainty in the image bone registration algorithm. Furthermore, the test studied images with MVCT slice thicknesses of 1 and 3 mm in unison with differing scan lengths. A sensitivity test was performed to determine the registration algorithm’s ability to find the absolute position of known translations/rotations of the pig head. The algorithm’s ability to determine absolute position was compared against that of manual operators, i.e. a radiation therapist and radiation oncologist. Finally, E2E tests for intracranial radiosurgery were performed by measuring the delivered dose distributions within the pig head using Gafchromic films.

Results: The repeatability test uncertainty was lowest for the MVCTs of 1-mm slice thickness, which measured less than 0.10 mm and 0.12 deg for all axes. For the sensitivity tests, the bone registration algorithm performed better than human eyes and a maximum difference of 0.3 mm and 0.4 deg was observed for the axes. E2E test results in absolute position difference measured 0.03 ± 0.21 mm in x-axis and 0.28 ± 0.18 mm in y-axis. A maximum difference of 0.32 and 0.66 mm was observed in x and y, respectively. The average peak dose difference between measured and calculated dose was 2.7 cGy or 0.4%.

Conclusion: Our tests using a pig head phantom estimate the TomoHDA system to have a submillimeter overall accuracy for intracranial radiosurgery.