A Phantom Design to Assist Patient Position Verification System in Daily Image-Guided RT and Comprehensive QA Measurements


J Syh

J Syh*, H Wu , Willis-Knighton Medical Center, Shreveport, LA

Presentations

SU-E-J-53 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:This study is to implement a homemade novel device with surface locking couch index to check daily radiograph (DR) function of adaPTInsight™, stereoscopic image guided system (SIGS), for proton therapy. The comprehensive daily QA checks of proton pencil beam output, field size, flatness and symmetry of spots and energy layers will be followed by using MatriXX dosimetry device.

Methods:The iBa MatriXX device was used to perform daily dosimetry which is also used to perform SIGS checks. A set of markers were attached to surface of MatriXX device in alignment of DRR of reconstructed CT images and daily DR. The novel device allows MatriXX to be fit into the cradle which was locked by couch index bars on couch surface. This will keep the MatriXX at same XY plane daily with exact coordinates. Couch height Z will be adjusted according to imaging to check isocenter-laser coincidence accuracy.

Results: adaPTInsight™ provides robotic couch to move in 6-degree coordinate system to align the dosimetry device to be within 1.0 mm / 1.0°. The daily constancy was tightened to be ± 0.5 mm / 0.3° compared to 1.0 mm / 1.0° before. For gantry at 0° and couch all 0° angles (@ Rt ARM 0 setting), offsets measured of the couch systems were ≤ 0.5° in roll, yaw and pitch dimensions.

Conclusion: Simplicity of novel device made daily image guided QA consistent with accuracy. The offset of the MatriXX isocenter-laser coincident was reproducible. Such easy task not only speeds up the setup, but it increases confidence level in detailed daily comprehensive measurements. The total SIGS alignment time has been shortened with less setup error. This device will enhance our experiences for the future QA when cone beam CT imaging modality becomes available at proton therapy center.