Evaluation of Deformable Image Registration Methods Between MRI and CT for Prostate Cancer Radiotherapy

no image available
N Wen

N Wen*, C Glide-Hurst , H Zhong , K Chin , A Kumarasiri , C Liu , M Liu , S Siddiqui , I Chetty , Henry Ford Health System, Detroit, MI

Presentations

SU-C-17A-3 Sunday 1:00PM - 1:55PM Room: 17A

Purpose: We evaluated the performance of two commercially available and one open source B-Spline deformable image registration (DIR) algorithms between T2-weighted MRI and treatment planning CT using the DICE indices.

Methods: CT simulation (CT-SIM) and MR simulation (MR-SIM) for four prostate cancer patients were conducted on the same day using the same setup and immobilization devices. CT images (120 kVp, 500 mAs, voxel size = 1.1x1.1x3.0 mm3) were acquired using an open-bore CT scanner. T2-weighted Turbo Spine Echo (T2W-TSE) images (TE/TR/α = 80/4560 ms/90°, voxel size = 0.7x0.7x2.5 mm3) were scanned on a 1.0T high field open MR-SIM. Prostates, seminal vesicles, rectum and bladders were delineated on both T2W-TSE and CT images by the attending physician. T2W-TSE images were registered to CT images using three DIR algorithms, SmartAdapt (Varian), Velocity AI (Velocity) and Elastix (Klein et al 2010) and contours were propagated. DIR results were evaluated quantitatively or qualitatively by image comparison and calculating organ DICE indices.

Results: Significant differences in the contours of prostate and seminal vesicles were observed between MR and CT. On average, volume changes of the propagated contours were 5%, 2%, 160% and 8% for the prostate, seminal vesicles, bladder and rectum respectively. Corresponding mean DICE indices were 0.7, 0.5, 0.8, and 0.7. The intraclass correlation coefficient (ICC) was 0.9 among three algorithms for the Dice indices.

Conclusion: Three DIR algorithms for CT/MR registration yielded similar results for organ propagation. Due to the different soft tissue contrasts between MRI and CT, organ delineation of prostate and SVs varied significantly, thus efforts to develop other DIR evaluation metrics are warranted.

Funding Support, Disclosures, and Conflict of Interest: Conflict of interest: Submitting institution has research agreements with Varian Medical System and Philips Healthcare.