Program Information
Radiation Dose to Patients Resulting From Image Guidance Procedures and AAPM TG-180 Update
G Ding1*, P Alaei2 , (1) Vanderbilt University , Nashville, TN, (2) Univ Minnesota, Minneapolis, MN
Presentations
SU-E-J-204 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose:
Image-guided radiation therapy (IGRT) is the new paradigm for patient positioning and target localization in radiotherapy. Daily imaging procedures add additional dose to the patient's treatment volume and normal tissues and may expose the organs at risk to unaccounted doses. This presentation is to update the progress of AAPM TG-180 which aims to provide strategies to quantify and account the dose from both MV and kV imaging in patient treatment planning.
Methods:
Our current knowledge on image guidance dose is presented. A summary of doses from image guidance procedures delivered to patients in relationship with therapeutic doses is given. Different techniques in reducing the image guidance dose are summarized. Typical organ doses resulting from different image acquisition procedures used in IGRT are tabulated.
Results:
Many techniques to reduce the imaging doses are available in clinical applications. There are large variations between dose to bone and dose to soft tissues for x-rays at kilovoltage energy range. Methods for clinical implementation of accounting for the imaging dose from an imaging procedure are available. Beam data from imaging systems can be generated by combining Monte Carlo simulations and experimental measurements for commissioning imaging beams in the treatment planning.
Conclusion:
The current treatment planning systems are not yet equipped to perform patient specific dose calculations resulting from kV imaging procedures. The imaging dose from current kV image devices has been significantly reduced and is generally much less than that resulting from MV. Because the magnitude of kV imaging dose is significantly low and the variation between patients is modest, it is feasible to estimate dose based on imaging producers or protocols using tabulated values which provides an alternative to accomplish the task of accounting and reporting imaging doses.
Contact Email: