Direct Realization of the CT Dose to Phantom: Energy to Heat Conversion in Polyethylene Using Calorimetry

no image available
H Chen-Mayer

H Chen-Mayer*, R Tosh , NIST, Gaithersburg, MD

Presentations

MO-E-17A-12 Monday 1:45PM - 3:45PM Room: 17A

Purpose: To develop a primary reference standard for absorbed dose to phantom for medical CT dosimetry. CT dosimetry relies on the implementation of the CTDI standard based on air kerma. We are taking a step toward an absorbed dose to water standard by first investigating the dose in a solid phantom. By directly measuring the heat converted from the incident radiation, the absorbed dose in the phantom at a point can be assessed with primary methods without relying on indirect conversions.

Methods: The calorimeter contains two small thermistors embedded in a removable PE “core” inserted into the cylindrical HDPE phantom. A core made with polystyrene (PS) was also tested because of its purportedly negligible heat defect. Measurements were made using the two cores and with a calibrated ionization chamber in a CT beam. The air chamber values were converted to the dose to medium using appropriate stopping-power ratios from the literature, and then compared to the thermal data.

Results: The PS core data yielded a dose of 1.3 times (4-run average, 3% std. dev.) higher than the converted chamber value, whereas the PE core data were inexplicably higher. The possible systematic errors include 1) excess heat from the thermistors, 2) in PE the exothermic chemical reactions, 3) uncertainties of the specific heat capacities of the materials, 4) thermal drift, and 5) theoretical conversion of chamber values. Monte Carlo simulations and finite element heat transfer calculations were performed to address some of these issues. The general validity was assessed in a 6 MV photon beam with an entirely different calibration scheme.

Conclusion: This study demonstrates the feasibility but also revealed the difficulty in developing a new primary reference standard for absorbed dose to material for CT. Additional experimental and theoretical work is planned to achieve our goal.