Encrypted login | home

Program Information

Dosimetric Comparison Between Cone, HDMLC and MicroMLC for the Treatment of Trigeminal Neuralgia


C Venencia

N Vacca , L Caussa , M Filipuzzi , E Garrigo , C Venencia*, Instituto Privado de Radioterapia - Fundacion Marie Curie, Cordoba, Argentina

Presentations

SU-E-T-122 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: The purpose of this work was to evaluate the dosimetric characteristics of three collimation systems, 5mm circular cone (Brainlab) and square fields of 5mm with HDMLC (Varian) and microMLC Moduleaf, Siemens) for trigeminal neuralgia treatment.

Methods: A TPS Iplan v4.5 BrainLAB was used to do treatment plans for each collimations system in a square solid water phantom with isocenter at 5cm depth. Single field and treatment plan including 11 arcs with fix field and 100° gantry range was made for each collimation systems. EBT3 films were positioned at isocenter in a coronal plane to measured dose distribution for all geometries. Films were digitized with a Vidar DosimetryPro Red scanner with a resolution of 89dpi and RIT113v6.1 software was used for analysis. Penumbra region (80%-20%), FWHM and dose percentage at 5mm and 10mm from CAX were determined. All profiles were normalized at CAX.

Results: For single beam the penumbra (FWHM) was 1.5mm (5.3mm) for the cone, 1.9mm (5.5mm) for HDMLC and 1.8mm (5.4mm) for the microMLC. Dose percentage at 5mm was 6.9% for cone, 12.5% for HDMLC and 8.7% for the microMLC. For treatment plan the penumbra (FWHM) was 2.58mm (5.47mm) for the cone, 2.8mm (5.84mm) for HDMLC and 2.58mm (6.09mm) for the microMLC. Dose perecentage at 5mm was 13.1% for cone, 16.1% for HDMLC, 15.2% for the microMLC.

Conclusion: The cone has a dose falloff larger than the microMLC and HDMLC, by its reduced penumbra, this translates into better protection of surrounding healthy tissue, however, the microMLC and HDMLC have similar accuracy to cone.


Contact Email: