Dosimetric and Mechanical Characteristics of Elekta Infinity LINAC with Agility MLC


J Park

J Park*, Q Xu , J Xue , Y Zhai , L An , Y Chen , MD Anderson Cancer Center at Cooper, Camden, NJ

Presentations

SU-E-T-119 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:
Elekta Infinity is the one of the latest generation LINAC with unique features. Two Infinity LINACs are recently commissioned at our institution. The dosimetric and mechanical characteristics of the machines are presented.

Methods:
Both Infinity LINACs with Agility MLC (160 leaves with 0.5 cm leaf width) are configured with five electron energies (6, 9, 12, 15, and 18 MeV) and two photon energies (6 and 15 MV). One machine has additional photon energy (10 MV). The commissioning was performed by following the manufacturer’s specifications and AAPM TG recommendations. Beam data of both electron and photon beams are measured with scanning ion chambers and linear diode array. Machines are adjusted to have the dosimetrically equivalent characteristics.

Results:
The commissioning of mechanical and imaging system meets the tolerances by TG recommendations. The PDD₁₀ of various field sizes for 6 and 15 MV shows < 0.5% difference between two machines. For each electron beams, R₈₀ matches with < 0.4 mm difference. The symmetry and flatness agree within 0.8% and 0.9% differences for photon beams, respectively. For electron beams, the differences of the symmetry and flatness are within 1.2% and 0.8%, respectively. The mean inline penumbras for 6, 10, and 15 MV are respectively 5.1±0.24, 5.6±0.07, and 5.9±0.10 mm for 10x10 cm at 10 cm depth. The crossline penumbras are larger than inline penumbras by 2.2, 1.4, and 1.0 mm, respectively. The MLC transmission factor with interleaf leakage is 0.5 % for all photon energies.

Conclusion:
The dosimetric and mechanical characteristics of two Infinity LINACs show good agreements between them. Although the Elekta Infinity has been used in many institutions, the detailed characteristics of the machine have not been reported. This study provides invaluable information to understand the Infinity LINAC and to compare the quality of commissioning data for other LINACs.