Validation and Implementation of Model-Based Patient Specific Quality Assurance Using Mobius3D and MobiusFX in a Clinical Setting


P Galavis

P Galavis*, K Osterman , G Jozsef , S Becker , K Dewyngaert , NYU Medical Center, New York, NY

Presentations

SU-E-T-532 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:This work carries out the commissioning and validation of the Mobius3D and MobiusFX software tools, which can replace the time-consuming measurement-based patient specific quality assurance (PSQA).

Methods:The beam model supplied by Mobius3D was validated against a 21EX linac’s beam measured data. Complex patient (VMAT) plans using Eclipse treatment planning system (TPS) was used to test the consistency between Mobius3D (calculates dose using patient image and field data) and MobiusFx (calculates dose using treatment dynalog files). Dose difference and gamma analysis (3%/3mm) between Mobius3D and MobiusFx were used to assess treatment plan and treatment delivery consistency. An end-to-end test was performed to validate Mobius3D and MobiusFx against ion chamber measurements. Effect of the dosimetric leaf gap (DLG) on Mobius3D dose calculation was additionally investigated.

Results: Mobius3D beam model parameters matched within 1%-3% with our beam measured data. A comparison of Mobius3D and MobiusFx dose matrices for VMAT planned prostate cases showed (0.33±0.07)% mean dose difference with gamma values above 95%. The end-to-end test showed dose differences of 1% between Mobius3D and MobiusFx. Dependence of Mobius3D dose calculation upon DLG was explored by introducing a ±0.5 mm change in the default value for DLG. This change resulted in agreement differences above 2%

Conclusion:Use of reference beam data would appear to speed up commissioning process for the clinical implementation of Mobius3D. However, careful consideration is required when comparing the information provided by the software, since large dose variations can be seen when the proper parameters are not optimized. The plan and delivered dose were in good agreement; hence MobiusFx has the potential to significantly speed up the PSQA process and at the same time helps to verify treatment parameters that are not possible with measurement-based PSQA.