no image available
Z Su

Z Su*, J Wu , Z Li , M Mamalui-Hunter , University of Florida/Radiation Oncology, Jacksonville, FL

Presentations

SU-E-J-194 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To quantify the Vero linac dosimetric accuracy of the tumor dynamic tracking treatment using EBT3 film embedded in a 4D thorax phantom.

Methods: A dynamic thorax phantom with tissue equivalent materials and a film insert were used in this study. The thorax phantom was scanned in 4DCT mode with a viscoil embedded in its film insert composed of lung equivalent material. Dynamic tracking planning was performed using the 50% phase CT set with 5 conformal beams at gantry angles of 330, 15, 60, 105 and 150 degrees. Each field is a 3cm by 3cm square centered at viscoil since there was no solid mass target. Total 3 different 1-2cos4 motion profiles were used with varied motion magnitude and cycle frequency. Before treatment plan irradiation, a 4D motion model of the target was established using a series of acquired fluoroscopic images and infrared markers motion positions. During irradiation, fluoroscopic image monitoring viscoil motion was performed to verify model validity. The irradiated films were scanned and the dose maps were compared to the planned Monte Carlo dose distributions. Gamma analyses using 3%-3mm, 2%-3mm, 3%-2mm, 2%-2mm criteria were performed and presented.

Results: For each motion pattern, a 4D motion model was built successfully and the target tracking performance was verified with fluoroscopic monitoring of the viscoil motion and its model predicted locations. The film gamma analysis showed the average pass rates among the 3 motion profiles are 98.14%, 96.2%, 91.3% and 85.61% for 3%-3mm, 2%-3mm, 3%-2mm, 2%-2mm criteria.

Conclusion: Target dynamic tracking was performed using patient-like breathing patterns in a 4D thorax phantom with EBT3 film insert and a viscoil. There was excellent agreement between acquired and planned dose distributions for all three target motion patterns. This study performed end-to-end testing and verified the treatment accuracy of tumor dynamic tracking.