MAGIC: Multi-Acquisition Gain Image Correction for Mobile X-Ray Systems with Intrinsic Localization Crosshairs


Y Park

Y Park*, G Sharp , Massachusetts General Hospital, Boston, MA

Presentations

WE-E-18A-7 Wednesday 1:45PM - 3:45PM Room: 18A

Purpose:Gain calibration for X-ray imaging systems with movable flat panel detectors (FPD) and intrinsic crosshairs is a challenge due to the geometry dependence of the heel effect and crosshair artifact. This study aims to develop a gain correction method for such systems by implementing the multi-acquisition gain image correction (MAGIC) technique.

Methods:Raw flat-field images containing crosshair shadows and heel effect were acquired in 4 different FPD positions with fixed exposure parameters. The crosshair region was automatically detected and substituted with interpolated values from nearby exposed regions, generating a conventional single-image gain-map for each FPD position. Large kernel-based correction was applied to these images to correct the heel effect. A mask filter was used to invalidate the original cross-hair regions previously filled with the interpolated values. A final, seamless gain-map was created from the processed images by either the sequential filling (SF) or selective averaging (SA) techniques developed in this study. Quantitative evaluation was performed based on detective quantum efficiency improvement factor (DQEIF) for gain-corrected images using the conventional and proposed techniques.

Results:Qualitatively, the MAGIC technique was found to be more effective in eliminating crosshair artifacts compared to the conventional single-image method. The mean DQEIF over the range of frequencies from 0.5 to 3.5 mm-1 were 1.09±0.06, 2.46±0.32, and 3.34±0.36 in the crosshair-artifact region and 2.35±0.31, 2.33±0.31, and 3.09±0.34 in the normal region, for the conventional, MAGIC-SF, and MAGIC-SA techniques, respectively.

Conclusion:The introduced MAGIC technique is appropriate for gain calibration of an imaging system associated with a moving FPD and an intrinsic crosshair. The technique showed advantages over a conventional single image-based technique by successfully reducing residual crosshair artifacts, and higher image quality with respect to DQE.


Contact Email: