Validation and Clinical Experience of Eclipse Electron Monte Carlo Algorithm (EMC)


 Pokharel

S Pokharel1*, S Rana2 , (1) 21st Century Oncology, Fort Myers, FL, (2) Procure Proton Therapy Center, Oklahoma City, Oklahoma

Presentations

SU-E-T-344 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: The purpose of this study is to validate Eclipse Electron Monte Carlo (Algorithm for routine clinical uses.

Methods: The PTW inhomogeneity phantom (T40037) with different combination of heterogeneous slabs has been CT-scanned with Philips Brilliance 16 slice scanner. The phantom contains blocks of Rando Alderson materials mimicking lung, Polystyrene (Tissue), PTFE (Bone) and PMAA. The phantom has 30X30x2.5 cm base plate with 2cm recesses to insert inhomogeneity. The detector systems used in this study are diode, tlds and Gafchromic EBT2 films. The diode and tlds were included in CT scans. The CT sets are transferred to Eclipse treatment planning system. Several plans have been created with Eclipse Monte Carlo (EMC) algorithm 11.0.21. Measurements have been carried out in Varian TrueBeam machine for energy from 6-22mev.

Results: The measured and calculated doses agreed very well for tissue like media. The agreement was reasonably okay for the presence of lung inhomogeneity. The point dose agreement was within 3.5% and Gamma passing rate at 3%/3mm was greater than 93% except for 6Mev(85%). The disagreement can reach as high as 10% in the presence of bone inhomogeneity. This is due to eclipse reporting dose to the medium as opposed to the dose to the water as in conventional calculation engines.

Conclusion: Care must be taken when using Varian Eclipse EMC algorithm for dose calculation for routine clinical uses. The algorithm dose not report dose to water in which most of the clinical experiences are based on rather it just reports dose to medium directly. In the presence of inhomogeneity such as bone, the dose discrepancy can be as high as 10% or even more depending on the location of normalization point or volume. As Radiation oncology as an empirical science, care must be taken before using EMC reported monitor units for clinical uses.