The Impact of Dose Calculation Algorithm and Hounsfield Units Conversion Tables On Plan Dosimetry for Lung SBRT

no image available
L Kuo

L Kuo*, E Yorke , S Lim , J Mechalakos , A Rimner , Memorial Sloan-Kettering Cancer Center, New York, NY

Presentations

SU-F-BRD-15 Sunday 4:00PM - 6:00PM Room: Ballroom D

Purpose:
To assess dosimetric differences in IMRT lung stereotactic body radiotherapy (SBRT) plans calculated with Varian AAA and Acuros (AXB) and with vendor-supplied (V) versus in-house (IH) measured Hounsfield units (HU) to mass and HU to electron density conversion tables.

Methods:
In-house conversion tables were measured using Gammex 472 density-plug phantom. IMRT plans (6 MV, Varian TrueBeam, 6-9 coplanar fields) meeting departmental coverage and normal tissue constraints were retrospectively generated for 10 lung SBRT cases using Eclipse Vn 10.0.28 AAA with in-house tables (AAA/IH). Using these monitor units and MLC sequences, plans were recalculated with AAA and vendor tables (AAA/V) and with AXB with both tables (AXB/IH and AXB/V). Ratios to corresponding AAA/IH values were calculated for PTV D95, D01, D99, mean-dose, total and ipsilateral lung V20 and chestwall V30. Statistical significance of differences was judged by Wilcoxon Signed Rank Test (p<0.05).

Results:
For HU<-400 the vendor HU-mass density table was notably below the IH table. PTV D95 ratios to AAA/IH, averaged over all patients, are 0.963±0.073 (p=0.508), 0.914±0.126 (p=0.011), and 0.998±0.001 (p=0.005) for AXB/IH, AXB/V and AAA/V respectively. Total lung V20 ratios are 1.006±0.046 (p=0.386), 0.975±0.080 (p=0.514) and 0.998±0.002 (p=0.007); ipsilateral lung V20 ratios are 1.008±0.041(p=0.284), 0.977±0.076 (p=0.443), and 0.998±0.018 (p=0.005) for AXB/IH, AXB/V and AAA/V respectively. In 7 cases, ratios to AAA/IH were within ± 5% for all indices studied. For 3 cases characterized by very low lung density and small PTV (19.99±8.09 c.c.), PTV D95 ratio for AXB/V ranged from 67.4% to 85.9%, AXB/IH D95 ratio ranged from 81.6% to 93.4%; there were large differences in other studied indices .

Conclusion:
For AXB users, careful attention to HU conversion tables is important, as they can significantly impact AXB (but not AAA) lung SBRT plans. Algorithm selection is also important for small targets in low density lung.