The Design of a Risk Index Method for 3D Patient Specific QA

no image available
W Cho

W Cho1*, L Xing2 , T Suh3 , H Wu1 , (1) Seoul National University Hospital, Seoul, (2) Stanford University, Stanford, CA, (3) Catholic Univ Medical College, Seoul,

Presentations

SU-E-T-316 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:To suggest a new guidance for the evaluation of 3D patient specific QA, a structure-specific risk-index (RI) method was designed and implemented.

Methods: A new algorithm was designed to assign the score of Pass, Fail or Pass with Risk to all 3D voxels in each structure by improving a conventional Gamma Index (GI) algorithm, which implied the degree of the risk of under-dose to the treatment target or over-dose to the organ at risks (OAR). Structure-specific distance to agreement (DTOA), dose difference and minimum checkable dose were applied to the GI algorithm, and additional parameters such as dose gradient factor and dose limit of structures were used to the RI method. Maximum passing rate (PR) and minimum PR were designed and calculated for each structure with the RI method. 3D doses were acquired from a spine SBRT plan by simulating the shift of beam iso-center, and tested to show the feasibility of the suggested method.

Results:When the iso-center was shifted by 1 mm, 2 mm, and 3 mm, the PR of conventional GI method between shifted and non-shifted 3D doses were 99.9%, 97.4%, and 89.7% for PTV, 99.8%, 84.8%, and 63.2% for spinal cord, and 100%, 99.5%, 91.7% for right lung. The minimum PRs from the RI method were 98.9%, 96.9%, and 89.5% for PTV, and 96.1%, 79.3%, 57.5% for spinal cord, and 92.5%, 92.0%, 84.4% for right lung, respectively. The maximum PRs from the RI method were equal or less than the PRs from the conventional GI evaluation.

Conclusion:Designed 3D RI method showed more strict acceptance level than the conventional GI method, especially for OARs. The RI method is expected to give the degrees of risks in the delivered doses, as well as the degrees of agreements between calculated 3D doses and measured (or simulated) 3D doses.