A Motorized Stage for Fast and Accurate QA of Machine Isocenter
J Moore*, E Velarde , J Wong , Johns Hopkins University, Baltimore, MD
Presentations
SU-E-T-373 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose: Precision delivery of radiation dose relies on accurate knowledge of the machine isocenter under a variety of machine motions. This is typically determined by performing a Winston-Lutz test consisting of imaging a known object at multiple gantry/collimator/table angles and ensuring that the maximum offset is within specified tolerance. The first step in the Winston-Lutz test is careful placement of a ball bearing at the machine isocenter as determined by repeated imaging and shifting until accurate placement has been determined. Conventionally this is performed by adjusting a stage manually using vernier scales which carry the limitation that each adjustment must be done inside the treatment room with the risks of inaccurate adjustment of the scale and physical bumping of the table. It is proposed to use a motorized system controlled outside of the room to improve the required time and accuracy of these tests.
Methods: The three dimensional vernier scales are replaced by three motors with accuracy of 1 micron and a range of 25.4mm connected via USB to a computer in the control room. Software is designed which automatically detects the motors and assigns them to proper axes and allows for small shifts to be entered and performed. Input values match calculated offsets in magnitude and sign to reduce conversion errors. Speed of setup, number of iterations to setup, and accuracy of final placement are assessed.
Results: Automatic BB placement required 2.25 iterations and 13 minutes on average while manual placement required 3.76 iterations and 37.5 minutes. The average final XYZ offsets is 0.02cm, 0.01cm, 0.04cm for automatic setup and 0.04cm, 0.02cm, 0.04cm for manual setup.
Conclusion: Automatic placement decreased time and repeat iterations for setup while improving placement accuracy. Automatic placement greatly reduces the time required to perform QA.
Contact Email: