Program Information
Characterization of Plastic Scintillator Detectors for Radiation Therapy
H Liu*, H Lin , A Darafsheh , J Finlay , S Both , T Zhu , Univ Pennsylvania, Philadelphia, PA
Presentations
SU-E-T-553 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose: To characterize basic performance of plastic scintillator detectors (PSD) designed for dosimetry of radiation therapy.
Methods: The Exradin W1 Scintillator is a plastic scintillating fiber-based detector designed for highly accurate measurement of small radiotherapy fields used in patient plan verification and machine commissioning and QA procedures. The Cerenkov emissions were corrected using spectral separation. The optical signal was converted to electronic signal with a photodiode. We measured its dosimetry performance, including percentage depth dose, output factor, dose and dose rate linear response. We compared the dosimetry results with reference ion chamber measurements.
Results: The dosimetry results of PSD agree well with reference ion chamber measurements. For percentage depth dose, the differences between PSD and ion chamber results are on average 1.7±1.1% and 0.8±0.8% with a maximum of 3.5% and 2.7% for 6MV and 15MV beams, respectively. For the output factors, PSD measurements are within 2% from ion chamber results. The dose linear response is within 1% when dose is larger than 20 MU for both 6 MV and 15 MV. The dose rate linear response is within 1% for the entire dose rate used (100 MU/min to 600MU/min).
Conclusions: The current design of PSD is feasible for the dosimtry measurement in radiation therapy. This combination of PSD and photodiode system could be extended to multichannel array detection of dose distribution. It might as well be used as range verification in proton therapy.
Funding Support, Disclosures, and Conflict of Interest: The work is partially supported by: DOD (W81XWH-09-2-0174) and American Cancer Society (IRG-78-002-28)
Contact Email: