Program Information
Spot-Scanning Proton Therapy Patient-Specific Quality Assurance: Our Methodology and Results From 295 Treatment Plans
D Mackin1*, X Zhang1 , Y Li2 , H Li1 , R Wu1 , F Poenisch1 , K Suzuki , M Kerr1 , C Holmes1 , X Zhu1 , N Sahoo1 , M Gillin1 , (1) The Univ. of Texas MD Anderson Cancer Center, Houston, TX, (2) Varian Medical Systems, Houston, TX
Presentations
MO-F-19A-5 Monday 2:45PM - 3:45PM Room: 19APurpose:
To report on the patient-specific quality assurance (PSQA) results for 295 spot-scanning proton therapy treatment plans from the MD Anderson PTC-Houston. We show how the results differed by treatment site and how they were affected by the treatment plan optimization method and by a range shifter in the treatment field. We also discuss some causes of PSQA problems.
Methods:
The PSQA procedure, which is designed to verify both the accuracy of the treatment planning system’s (Eclipse™ v8.9) dose calculations and the dose delivery of the Hitachi PROBEAT synchrotron, consists of (1) an end-to-end test in which the beam is delivered and measured at the prescribed gantry angle, and (2) additional dose plane measurements made from gantry angle 270°. HPlusQA™ software automatically performs the gamma analysis with criteria 3% (dose tolerance), 3 mm (distance-to-agreement, DTA) and 2%, 2 mm. Passing is defined as at least 90% of the pixels having a gamma score less than 1.
Results:
The PSQA gamma passing rate was 96.2% for 3%, 3 mm, and 85.3% for 2%, 2 mm. The rate depended on the treatment site. For example, the 3%, 3 mm passing rate was 95% for head and neck plans, vs 100% for prostate plans. The passing rates of multi- vs. single-field optimization plans did not significantly differ. However, the rate for fields with range shifters was 94.8±0.6%, vs 99.0±0.6% for those without (p = 0.002). Longitudinal dose gradients caused most of the low scores. Overestimation of the calculated dose proximal to the spread-out Bragg peak (SOBP) caused many of the others.
Conclusion:
The planned and delivered doses consistently agreed within tolerance levels. Minor dose modeling deficiencies remain proximal to the SOBP. The 3% dose tolerance, 3 mm DTA, with 90% pixel passing rate is a reasonable action level for 2D gamma comparisons.
Contact Email: