Program Information
Practical Considerations On Implementation of the Image Gently Pediatric CT Protocols
J Zhang1*, E Woods2 , C Adams3 , C Lumby4 , E Richer5 , J Dillon6 , (1) University of Kentucky, Lexington, KY, (2) University of kentucky, Lexington, KY, (3) University of Kentucky, Lexington, KY, (4) University of Kentucky, Lexington, KY, (5) University of Kentucky, Lexington, Lexington, (6) University of Kentucky, Lexington, KY
Presentations
SU-E-I-68 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose: One limitation associated with the Image Gently pediatric CT protocols is practical implementation of the recommended manual techniques. Inconsistency as a result of different practice is a possibility among technologist. An additional concern is the added risk of data error that would result in over or underexposure. The Automatic Exposure Control (AEC) features automatically reduce radiation for children. However, they do not work efficiently for the patients of very small size and relative large size. This study aims to implement the Image Gently pediatric CT protocols in the practical setting while maintaining the use of AEC features for pediatric patients of varying size.
Methods:Anthropomorphological abdomen phantoms were scanned in a CT scanner using the Image Gently pediatric protocols, the AEC technique with a fixed adult baseline, and automatic protocols with various baselines. The baselines were adjusted corresponding to patient age, weight and posterior-anterior thickness to match the Image Gently pediatric CT manual techniques. CTDIvol was recorded for each examination. Image noise was measured and recorded for image quality comparison. Clinical images were evaluated by pediatric radiologists.
Results:By adjusting vendor default baselines used in the automatic techniques, radiation dose and image quality can match those of the Image Gently manual techniques. In practice, this can be achieved by dividing pediatric patients into three major groups for technologist reference: infant, small child, and large child. Further division can be done but will increase the number of CT protocols. For each group, AEC can efficiently adjust acquisition techniques for children. This implementation significantly overcomes the limitation of the Image Gently manual techniques.
Conclusion:Considering the effectiveness in clinical practice, Image Gently Pediatric CT protocols can be implemented in accordance with AEC techniques, with adjusted baselines, to achieve the goal of providing the most appropriate radiation dose for pediatric patients of varying sizes.
Contact Email: