Encrypted login | home

Program Information

An Improved Formula of Scatter-To-Primary Ratio for Photon Dose Calculation


T Zhu

T Zhu1*, (1) Univ Pennsylvania, Philadelphia, PA

Presentations

SU-E-T-439 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: Scatter-to-primary ratio (SPR) is an important dosimetric quantity that describes the contribution from the scatter photons in an external photon beam. The purpose of this study is to develop an improved analytical formula to describe SPR as a function of circular field size (r) and depth (d) using Monte Carlo (MC) simulation.

Methods: MC simulation was performed for Mohan photon spectra (Co-60, 4, 6, 10, 15, 23 MV) using EGSNRC code. Point-spread scatter dose kernels in water are generated. The scatter-to-primary ratio (SPR) is also calculated using MC simulation as a function of field size for circular field size with radius r and depth d. The doses from forward scatter and backscatter photons are calculated using a convolution of the point-spread scatter dose kernel and by accounting for scatter photons contributing to dose before (z’d) reaching the depth of interest, d, where z’ is the location of scatter photons, respectively. The depth dependence of the ratio of the forward scatter and backscatter doses is determined as a function of depth and field size.

Results: We are able to improve the existing 3-parameter (a, w, d0) empirical formula for SPR by introducing depth dependence for one of the parameter d0, which becomes 0 for deeper depths. The depth dependence of d0 can be directly calculated as a ratio of backscatter-to-forward scatter doses for otherwise the same field and depth. With the improved empirical formula, we can fit SPR for all megavoltage photon beams to within 2%. Existing 3-parameter formula cannot fit SPR data for Co-60 to better than 3.1%.

Conclusion: An improved empirical formula is developed to fit SPR for all megavoltage photon energies to within 2%.


Contact Email: