Encrypted login | home

Program Information

Improvement of TomoTherapy Megavoltage Topogram Image Quality for Automatic Registration During Patient Localization

no image available
J Scholey

J Scholey1*, S Qi2 , D Low2 , B White1 , (1) University of Pennsylvania, Philadelphia, PA, (2) UCLA School of Medicine, Los Angeles, CA

Presentations

SU-D-BRF-3 Sunday 2:05PM - 3:00PM Room: Ballroom F

Purpose: To improve the quality of mega-voltage orthogonal scout images (MV topograms) for a fast and low-dose alternative technique for patient localization on the TomoTherapy HiART system.

Methods: Digitally reconstructed radiographs (DRR) of anthropomorphic head and pelvis phantoms were synthesized from kVCT under TomoTherapy geometry (kV-DRR). Lateral (LAT) and anterior-posterior (AP) aligned topograms were acquired with couch speeds of 1cm/s, 2cm/s, and 3cm/s. The phantoms were rigidly translated in all spatial directions with known offsets in increments of 5mm, 10mm, and 15mm to simulate daily positioning errors. The contrast of the MV topograms was automatically adjusted based on the image intensity characteristics. A low-pass fast Fourier transform filter removed high-frequency noise and a Weiner filter reduced stochastic noise caused by scattered radiation to the detector array. An intensity-based image registration algorithm was used to register the MV topograms to a corresponding kV-DRR by minimizing the mean square error between corresponding pixel intensities. The registration accuracy was assessed by comparing the normalized cross correlation coefficients (NCC) between the registered topograms and the kV-DRR. The applied phantom offsets were determined by registering the MV topograms with the kV-DRR and recovering the spatial translation of the MV topograms.

Results: The automatic registration technique provided millimeter accuracy and was robust for the deformed MV topograms for three tested couch speeds. The lowest average NCC for all AP and LAT MV topograms was 0.96 for the head phantom and 0.93 for the pelvis phantom. The offsets were recovered to within 1.6mm and 6.5mm for the processed and the original MV topograms respectively.

Conclusion: Automatic registration of the processed MV topograms to a corresponding kV-DRR recovered simulated daily positioning errors that were accurate to the order of a millimeter. These results suggest the clinical use of MV topograms as a promising alternative to MVCT patient alignment.


Contact Email: