Dosimetric Evaluation of a Novel CT/MR Compatible Fletcher Applicator for Intracavitary Brachytherapy of the Cervix Uteri

no image available
K Gifford

K Gifford1*, T Han2 , F Mourtada3 , P Eifel4 , (1) UT MD Anderson Cancer Center, Houston, TX, (2) UT MD Anderson Cancer Center, Houston, TX, (3) Christiana Care Hospital, Newark, DE, (4) The UT MD Anderson Cancer Center, Houston, Texas

Presentations

SU-F-19A-4 Sunday 4:00PM - 6:00PM Room: 19A

Purpose:To validate a Monte Carlo model and evaluate the dosimetric capabilities of a novel commercial CT/MR compatible Fletcher applicator for cervical cancer brachytherapy.

Methods:MCNPX 2.7.0 was used to model the Fletcher CT/MR shielded applicator (FA) and 192Ir HDR source. Energy deposition was calculated with a track length estimator modified by an energy-dependent heating function. A high density polystyrene phantom was constructed with three film pockets for validation of the MCNPX model. Three planes of data were calculated with the MCNPX model corresponding to the three film planes in phantom. The planes were located 1 cm from the most anterior, posterior, and medial extents of the FA right ovoid. Unshielded distributions were calculated by modeling the shielded cells as air instead of the tungsten alloy. A third order polynomial fit to the OD to dose curve was used to convert OD of the three film planes to dose. Each film and MCNPX plane dose distribution was normalized to a point 2 cm from the center of the film plane and in a region of low dose gradient. MCNPX and film were overlaid and compared with a distance-to-agreement criterion of (±2%/±2mm). Shielded and unshielded distributions were overlaid and a percent shielded plot was created.

Results:85.2%, 97.1%, and 96.6% of the MCNPX points passed the (±2%/±2mm) criterion respectively for the anterior, lateral, and posterior film comparison planes. A majority of the points in the anterior plane that exceeded the DTA criterion were either along edges of where the film was cut or near the terminal edges of the film. The percent shielded matrices indicated that the maximum % shielding was 50%.

Conclusion:These data confirm the validity of the FA Monte Carlo model. The FA ovoid can shield up to 50% of the dose in the anteroposterior direction.