Program Information
Evaluation of the Lens Dose On the Cone Beam IGRT Procedures
R Palomo-Llinares1 , M Pujades2 , J Gimeno-Olmos1 , V Carmona Meseguer1 , F Lliso-Valverde1 , C Candela-Juan1 , F Ballester7 , J Perez-Calatayud1*, (1) Hospital La Fe, Valencia, Valencia, (2) National Dosimetry Center, Valencia, Valencia, (3) University of Valencia, Burjassot
Presentations
SU-E-J-69 Sunday 3:00PM - 6:00PM Room: Exhibit HallPurpose: With the establishment of the IGRT as a standard technique, the extra dose that is given to the patients should be taken into account. Furthermore, it has been a recent decrease of the dose threshold in the lens, reduced to 0.5 Gy (ICRP ref 4825-3093-1464 on 21st April, 2011).
The purpose of this work was to evaluate the extra dose that the lens is receive due to the Cone-Beam (CBCT) location systems in Head-and-Neck treatments.
Methods: The On-Board Imaging (OBI) v 1.5 of the two Varian accelerators, one Clinac iX and one True Beam, were used to obtain the dose that this OBI version give to the lens in the Head-and-Neck location treatments. All CBCT scans were acquired with the Standard Dose Head protocol (100 kVp, 80 mA, 8 ms and 200 degree of rotation).
The measurements were taken with thermoluminescence (TLD) EXTRAD (Harshaw) dosimeters placed in an anthropomorphic phantom over the eye and under 3 mm of bolus material to mimic the lens position. The center of the head was placed at the isocenter. To reduce TLD energy dependence, they were calibrated at the used beam quality.
Results: The average lens dose at the lens in the OBI v 1.5 systems of the Clinac iX and the True Beam is 0.071 and 0.076 cGy/CBCT, respectively.
Conclusions: The extra absorbed doses that receive the eye lenses due to one CBCT acquisition with the studied protocol is far below the new ICRP recommended threshold for the lens. However, the addition effect of several CBCT acquisition during the whole treatment should be taken into account.
Contact Email: