Program Information
Assessing Elekta MLC Tracking Performance
M Glitzner*, S Crijns , J Lagendijk , B Raaymakers , University Medical Center Utrecht, Utrecht, The Netherlands
Presentations
SU-D-18A-5 Sunday 2:05PM - 3:00PM Room: 18APurpose:
To investigate the technical feasibility of using Elekta MLCs in tracked radiotherapy, specifically to show the capability of an Agility MLC to follow model tumor trajectories using object tracking. The system performance of the MLC is to be assessed via dosimetric experiments and temporal response measures.
Methods:
The system observer in the tracking cascade consisted of a camera serving for this proof of concept; Images of a moving object were acquired continuously and decorrelated for the position information, which was fed forward to the RT computer controlling the MLC via an interface provided by Elekta. The 1D motion (in the direction of the leaves) was induced by a CIRS motion controller. Additionally, the aperture position was observed by means of the light field of the linac. Calculating the phase between aperture and object yielded the mean aperture lag for the current setup. The object was restricted to sinusoidal motion with a period of 10s. For dosimetry, GafChromic radiosensitive film was irradiated with a total dose of 1000MU using an aperture size of
40mm and a motion range of 30mm.
Results:
Object tracking can substantially reconstruct the geometric dose response of a static target. In the dynamic case, dose is smeared out into the legs of the static distribution, leading to a reduced plateau and increased FWHM of 5mm, compared to the static width of 38mm. The time-lag between object and aperture was determined to be approximately 300ms for the current set-up.
Conclusion:
We demonstrated a tracking experiment performed on a clinical Elekta linear accelator for the first time. Observed profile variations show the dosimetric impact of tracked delivery. The determined lag is a valuable descriptor for a future tracking cascade employing predictor filters. The performed experiments are generic and possible predecessors for future applications with MR-Linac or ultrasound probes.
Funding Support, Disclosures, and Conflict of Interest: Conflict of interest: this project is partly funded by Elekta.